The Machinery Works

Home > Ophthalmic Hints >
Marco Lensmeter with separated targets

Here's a simple fix that you can apply to fix Marco lensmeter targets that separate. Remove the target cell from the instrument. Remove the Risley prism assembly also. Turn the instrument so it is pointing straight up. Put the target back in the instrumen